상품명 | KUV-6000T Touch Screen UV/Visible Spectrophotometer |
---|---|
판매가 | 가격문의 |
수량 |
SNS 상품홍보 | |
---|
위 옵션선택 박스를 선택하시면 아래에 상품이 추가됩니다.
상품명 | 상품수 | 가격 |
---|
Notice
- 주문하시기전에 재고 여부 및 배송일정에 대한 상담을 먼저 진행해 주시기 바랍니다. 02-6264-5745
- Process QuestionNaire 를 작성해 주시면 검토 후 제품을 추천해 드리겠습니다. [양식다운로드]
- 독일 IKA 재고상품을 특별가격으로 구입하실 수 있습니다. [기획전 바로가기]
제품상세정보
제품명 : KUV-6000T Touch Screen UV/Visible Spectrophotometer
제품특징
• Intelligent Alarm function
• System Encryption function
• Display several groups of data at the same time for direct reading
• Module design for easy operation
• New panel with touch screen
• 800*480 touch screen displays which shows clear graphs
• Stand-alone system which can realize spectrum measurement, quantitative, wavelength scanning, kinetics, system setting directly on the instrument
• The instrument uses a rigid 8mm (KUV-6000T) or 16mm (KUV-8000T) die-cast aluminum base as its optical mount to ensure instrument stability and reliability
• Automatic system calibration and positioning to recover deviation after long-time use
• Imported main optical spare parts to ensure low stray light, stability and reliability
• The instrument has powerful memory which can print data and graphs in A4 format directly by connecting to a printer
• Data can export to USB directly
제품사양
Model | KUV-6000T | Photometric Range | 0-200%T, -4.0-4.0A, 0-9999C |
---|---|---|---|
Wavelength Range | 190-1100nm | Baseline flatness | ±0.001A |
Spectral Bandwidth | 2nm | Noise | ±0.0005A |
Wavelength Accuracy | ±0.5nm | Display | 800*480 Smart touch screen |
Wavelength Repeatability | <0.2nm | Data Output Por | USB |
Stray Light | <0.05%T220nm,360nm | Print Port | USB Port |
Photometric Accuracy | ±0.2T(0-100%T) ±0.002Abs(0-0.5Abs) ±0.004Abs(0.5-1.0Abs) |
Light source : Tungsten Lamp&D2 lamp |
Power Requirements : AC220V/50Hz orl 10V/60Hz |
Photometric Repeatability | <0.05%T(0-100%T) 0.001Abs(0-0.5Abs) 0.002Abs(0.5-1.0Abs) |
Dimension | 625*430*206mm |
Stray Light | <0.03%T @220nm,360nm | Detector : Silicon Photodiode | Dimension : 460*380*180nm |
Stability | ±0.001A/h@500mn | Weight | 20kg |
Model | KUV-8000T | Photometric Range | 0-200%T, -4.0-4.0A, 0-9999C |
---|---|---|---|
Wavelength Range | 190-1100nm | Baseline flatness | ±0.001A |
Spectral Bandwidth | 1.8nm | Noise | ±0.0004A |
Wavelength Accuracy | ±0.1nm(D2 656. lnm), ±0.3nm full range | Display | 800*480 Smart touch screen |
Wavelength Repeatability | <0.1nm | Data Output Por | USB |
Stray Light | <0.03%T220nm,360nm | Print Port | USB Port |
Photometric Accuracy | ±0.2T(0-100%T) ±0.002Abs(0-0.5Abs) ±0.004Abs(0.5-1.0Abs) |
Light source : Tungsten Lamp&D2 lamp |
Power Requirements : AC220V/50Hz or 110V/60Hz |
Photometric Repeatability | <0.05%T(0-100%T) 0.001Abs(0-0.5Abs) 0.002Abs(0.5-1.0Abs) |
Detector | Silicon Photodiode |
Stability | ±0.0004A/h@500mn | Weight | 28kg |